Vacuum drying oven VDL023-230V-EU

BINDER

Type Nominal capacity l Max. grids Included grids Internal dimensions (W x D x H) mm External dimensions (W x D x H) mm Rated capacity kW Type of auxiliary energy PK
Purchase order number
4678320
Cat. No. Type Nominal capacity l Max. grids Included grids Internal dimensions (W x D x H) mm External dimensions (W x D x H) mm Rated capacity kW Type of auxiliary energy PK

1

4678317 VD 23 24 4 1 285 x 295 x 285 523 x 413 x 698 0.9 200 ... 230 V, 50/60 Hz 1
4678318 VD 56 55 5 1 400 x 343 x 400 638 x 461 x 815 1.4 200 ... 230 V, 50/60 Hz 1
4678319 VD 115 119 6 1 506 x 460 x 506 743 x 581 x 942 1.6 200 ... 230 V, 50/60 Hz 1
4678320 VDL 23 24 4 2 285 x 295 x 285 523 x 413 x 698 0.9 200 ... 230 V, 50/60 Hz 1
4678321 VDL 56 55 5 2 400 x 343 x 400 638 x 461 x 815 1.4 200 ... 230 V, 50/60 Hz 1
4678322 VDL 115 119 6 2 506 x 460 x 506 743 x 581 x 942 1.6 200 ... 230 V, 50/60 Hz 1

Für schonende, schnelle Trocknung ohne Rückstände, Verkrustungen oder Oxidation.

  • Direkte Wärmeübertragung durch große Wärmeleitplatten
  • Programmgesteuerte Trocknungsüberwachung mit automatischer Belüftung bei Prozessende
  • Interner Datenlogger, Messwerte im offenen Format über USB auslesbar
  • Universalbelüftung, für Inertgas oder Umgebungsluft nutzbar
  • Sicherheitsglasscheibe, federnd gelagert mit Splitterschutz
  • Großes Sichtfenster
  • Spanneinschübe aus Aluminium, individuell positionierbar
  • Ethernet-Schnittstelle

Vakuumtrockenschränke VD:
  • Temperaturbereich: RT +10 °C bis 220 °C
  • Für nicht entflammbare Lösungsmittel
  • Controller mit digitaler Anzeige von Druck und Temperatur
  • 2 Schaltausgänge 24 V DC (max. 0,4 A)

Vakuumtrockenschränke VDL:
  • Temperaturbereich: RT +10 °C bis 110 °C
  • Für entflammbare Lösungsmittel
  • ATEX-Konformität Geräte: EX II 2/3/- G IIB T3 Gb/Gc/- X
  • Intuitiver Touchscreen Controller mit graphischer Anzeige von Druck und Temperatur
  • Druckwächter für Heizungsfreigabe ab < 100 mbar