| Cat. No. | Type | Nominal capacity l | Max. grids | Internal dimensions (W x D x H) mm | External dimensions (W x D x H) mm | Type of auxiliary energy | Rated capacity kW | Weight kg | PK | |
|---|---|---|---|---|---|---|---|---|---|---|
| 4685704 | MKF 400 | 417 | 6 | 735 x 810 x 700 | 1115 x 1400 x 1710 | 400 V, 50 Hz | 8 | 432 | 1 | |
| 4685705 | MKF 400 | 417 | 6 | 735 x 810 x 700 | 1115 x 1400 x 1710 | 480 V, 60 Hz | 8 | 432 | 1 | |
| 4685706 | MKF 1020 | 991 | 11 | 1200 x 810 x 1020 | 1580 x 1480 x 2005 | 400 V, 50 Hz | 12.5 | 636 | 1 | |
| 4685707 | MKF 1020 | 991 | 11 | 1200 x 810 x 1020 | 1580 x 1480 x 2005 | 480 V, 60 Hz | 12.5 | 636 | 1 | |
| 7984336 | MKF 56 | 60 | 4 | 400 x 348 x 420 | 720 x 780 x 1445 | 200 ... 230 V, 50 Hz | 2.8 | 175 | 1 | |
| 7984339 | MKF 240 | 228 | 6 | 735 x 443 x 700 | 1115 x 925 x 1715 | 400 V, 50 Hz | 6.8 | 360 | 1 |
Environmental chambers are ideal for material testing under dynamic conditions. The standard rapid alternation speeds of approx. 5 K/min enable standard-compliant testing.
- Homogeneous climate conditions thanks to APT.line™ preheating chamber technology
- Touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out via USB
- Programmable condensation protection
- Troubleshooting system with visual and acoustic alarms
- CFC-free refrigerant R-452A
- Heated viewing window with LED interior lighting
- 1 stainless steel rack
-Humidity regulation with capacitive humidity sensor and responsive steam humidification
- Automatic water and wastewater management
- Integrated 20 l water storage tank
- 4 zero-voltage relay contacts
Models with 480 V nominal voltage are equipped with a voltage and frequency converter.
Accessories
Alternatives